Shanghai Chemtron Biotech Co.Ltd.

Report of Chemtron H.P.Antibody Detection Test(golden colloid)

Abstract:
Clinical evaluation is made on H.P Antibody Detection Test (golden colloid) manufactured by Chemtron Biotech Inc. and 1034 cases are analysed . Compared with 14 C-UBT,the sensitivity is 92.9%,specification is 96.2% and accordance rate is 94.3%.The accordance rate between the two detection methods is very high.The test is non-invasive,rapid ,simple and of high sensitivity and specification. It can be use as an aid in the diagnosis of H.P infection.


1.Introduction

Helicobater pylori are gram negative bacteria that infect gastric mucosa. Infection with Helicobacter pylori may lead to chronic gastritis and predispose to gastric and duodenal ulcers. Infection with H.pylori is very common and has been estimated to occur in 40-50%? of? the population in developed countries and 80-90% of the population in developing countries. Now H.P. has been listed on the first catalogue of cancer-resulting factors. The early detection of H.P infection is useful for the early treatment and prevent the further damage to the human body.?
Several techniques, both invasive and non-invasive have been? utilized? to? diagnose? H.pylori? infection.? The urea breath test is based on the detection of labeled carbon dioxide in expired air as a result or urease production by H.pylori. Serological tests are based on the detection of a specific anti-H.pylori antibody, mostly by IgG antibodies in a patient’s blood.?
Different invasive techniques are used for the diagnosis of H.pylori infection. One of which is the endoscopic biopsy of? antrum? or? corpus? of? stomach? to? detect? H.pylori? by specialized stains such as Giemsa, culture and rapid urease test.? These? methods? need special? laboratory? equipment and trained technicians. Also, the culture method takes a long? time? to? produce? a? result.? Serological? approaches detect the host’s immune response to the infection, but do not permit a doctor to distinguish a current active infection from a past infection. In addition, the antibody titer does not change during the treatment. The? 14 C-UBT and 13 C-UBT and N15 emission test is with a relatively high cost ,a restriction by specialized reagents and particular detection equipment but a accurate results. It’s regarded as the golden standard of the Helicobater pylori detection.
The? Chemtrue? ?H.pylori? Antibody? Rapid? Test? is? a chromatographic? immunoassay? for? the? qualitative determination of anti-H.pylori. IgG antibodies in human serum, plasma and whole blood .It can be used extensively in the initial screening or as an aid in the diagnosis of H.P infection and the examination in the prognosis.


2.Purpose for evaluation

To evaluate the clinical performance of Chemtron H.P antibody test such as? sensitivity ,specification and accuracy and etc.


3.The design of Clinical trial

3.1 The samples of clinical trial
3.2 There are 1034 cases in this trial ,consisted of 706 male cases and 328 female cases, aging from 15-82 .
3.3 Materials of trial?
3.3.1Chemtron HPAb test (golden colloidal) ,batach No.20060402,20060703
3.3.214 C-UBT test: manufactured by China nuclear –dewei biotech ltd. National Medicine Licence No.S20000021,batch No.20060501,20080301
3.3 Method of evaluation:
3.3.1 Every sample is tested by Chemtron H.P.test and 14 C-UBT respectively and make a comparison.
3.3.2 Test procedure: Operate according to the product instruction.


4.Results of clinical trial

Results of tests on 1034 cases compared with 14 C-UBT
14 C-UBT positive 590,negative 444;Chemtron H.P. antibody test (golden colloidal) positive 565,negative 469;Positive in common 548,negative in common 427.
Compared with 14 C-UBT,Chemtron H.P. antibody Test(golden colloid ):
Sensitivity=548/(548+42) * 100%=92.9%
Specification=427/(17+427)*100%=96.2%
Accordence rate=(548+427)/(548+17+42+427)*100%=94.3%
Predicting value of postive tests=548/(548+17)*100%=97.0%
Predicting value of negative tests=427/(42+427)*100%=91.0%

5. Conclusion
Results of clinical tests shows:Compared with 14 C-UBT test,the sensitivity of Chemtron H.P.Antibody test is 92.9%,specification is 96.2% and accordance rate is 94.3%.It has a high rate of accordance and a good clinical performance.The test procedure is rapid and simple and can be used as an aid in the detection of H.P. infection.?
Related News
  • Menopause Test

    Menopause Test

    March 30, 2018Menopause is the period of time when every woman goes from youth to an old age, which is like a "threshold".Since the age of 40, menopausal symptoms have been troubling women, including facial skin fl...view
  • Clinical Report for Chlamydia Antigen Rapid Test Kit

    Clinical Report for Chlamydia Antigen Rapid Test Kit

    March 30, 2018Chlamydia infection is one of the most common sexually transmitted diseases, the main damage is to cause cervicitis, nonspecific urethritis. Up to now, Chlamydia infection is increasing every year in ...view
  • Drug Testing: Classification Of Drug

    Drug Testing: Classification Of Drug

    March 30, 2018Classification of drugAccording to the IOC classification, drug mainly has the following seven categories: stimulants, synthetic steroids, narcotic analgesics, diuretics, blood stimulants, peptides an...view
  • At Home Drug Test

    At Home Drug Test

    March 30, 2018Harm of drug abuse to the society1. The harm to the family: Once there is a drug user in the family, the family will not be peaceful. The drug addicts destroy themselves while damaging their family, c...view
  • The Importance Of Hormone Test

    The Importance Of Hormone Test

    March 30, 2018Hormones is originated from the Greek, that is incretion, meaning "activated." Later scholars defined it as chemical substances which produce by endocrine organ , and then released into the blood circ...view
  • New Fertility Kits

    New Fertility Kits

    March 30, 2018Roche recently announced the launch of a new blood test kit called Elecsys AMH (anti-mullerian hormone). The new fertility test kit evaluate female’s ovarian reserve function according to indication ...view
  • TEL:+86-21-68129666
  • FAX:+86-21-68129222
  • EMAIL:[email protected]
  • ADDRESS:No.518, Qingdai Rd., International Medical Park, Pudong, 201318, Shanghai, P.R.China
北单500奖金计算器